What a single-quadrupole instrument can honestly confirm.

first seen 2024-Q30 records submitted264 annotations
Section 1

Contribution

Comments technically on records submitted by others. Annotations are signed and are not moderated for agreement.

Records submitted 0 n=0
Annotations 264 n=264
Median purity submitted suppressed n=0 n<3 — see individual records
Suppliers covered 0 n=0

This handle has submitted no records. Annotators comment on other people's submissions and are not required to submit.

Section 2

Annotations

Signed technical comments on records. The index does not moderate annotations for agreement with its own arithmetic, and several of these disagree with it. Where a handle has made the same point on several records, one instance is shown and the rest are reachable from the records themselves.

AX-2026-041867
Instrument class matters here. A unit-resolution determination reported to a tenth of a ppm is reporting a precision the hardware does not have, and -36.5 ppm on such a determination is largely noise.
AX-2026-041599
Identity confirmed and purity at 97.96 %. The two answer different questions — what it is, and how much else is in it — and a record carrying both is the version worth citing.
AX-2026-043295
Instrument class is on the record, which is what allows -2.2 ppm to be read at its proper weight. The index distinguishes "confirmed" from "consistent with" precisely on this field.
AX-2026-043440
Instrument class is on the record, which is what allows -3.4 ppm to be read at its proper weight. The index distinguishes "confirmed" from "consistent with" precisely on this field.
AX-2026-043327
Instrument class is on the record, which is what allows -4.4 ppm to be read at its proper weight. The index distinguishes "confirmed" from "consistent with" precisely on this field.
AX-2026-043166
The index grades identity by instrument class, not by the ppm figure alone: "confirmed" where the determination was made at high resolution, "consistent with" where it was not. The status block above says which applies to this record, and it is worth reading before the number.
AX-2026-042127
A unit-resolution instrument confirms that a species of approximately the right mass is present. It never confirms a species of exactly the right mass, and the index's status vocabulary distinguishes the two deliberately.
AX-2026-042220
Instrument class is on the record, which is what allows +1.2 ppm to be read at its proper weight. The index distinguishes "confirmed" from "consistent with" precisely on this field.
AX-2026-043470
Instrument class is on the record, which is what allows +2.0 ppm to be read at its proper weight. The index distinguishes "confirmed" from "consistent with" precisely on this field.
AX-2026-041973
Instrument class is on the record, which is what allows -1.6 ppm to be read at its proper weight. The index distinguishes "confirmed" from "consistent with" precisely on this field.
AX-2026-041331
The index records this as identity not established. That is a statement about the evidence, not about the material, and the distinction is the whole reason the status vocabulary has four values instead of two.
AX-2026-041971
Instrument class is on the record, which is what allows +3.4 ppm to be read at its proper weight. The index distinguishes "confirmed" from "consistent with" precisely on this field.

Showing 12 distinct comments of 264 annotations by this handle.