AX-2024-004719

Oxytocin · 2 mg declared · Hunan Aslsen Technology Co., Ltd. · lot HOMO-2405-U3389

MARGINALIDENTITY CONSISTENT
Report ID
AX-2024-004719
Compound
Oxytocin
Compound class
Hypothalamic and posterior-pituitary hormones
Supplier
Homopeptide — Hunan Aslsen Technology Co., Ltd.
Batch / lot
HOMO-2405-U3389
Declared label claim
2mg
Laboratory
Medutest
Assay date
13 Sep 2024
Submitted
17 Sep 2024
Quarter
2024-Q3
Submitter
e.almeida
Sample provenance
Buyer-sourced
Section 1

Results

Measured against the index acceptance specification for the hypothalamic and posterior-pituitary hormones class. A specification is a decision the index made; a result is a measurement. See where each threshold came from.

ParameterMethodResultUnitIndex spec
PurityRP-HPLC, 25 min gradient, 214 nm99.73% area≥ 98.5PASS
Total impuritiesRP-HPLC, sum of reported peaks0.27% area≤ 1.5PASS
Largest single impurityRP-HPLC — open-chain (reduced disulfide, +2 Da)0.10% area≤ 1.00PASS
Identity — theoretical monoisotopic masscalculated from molecular formula1006.4365DaNOT TESTED
Identity — observed monoisotopic massLC-MS, unit resolution (single quadrupole)1006.4364DaNOT TESTED
Mass accuracyobserved vs theoretical-0.1ppm± 15PASS
Peptide contentnot ordered on this submissionnot determinedrequired to check label claimNOT TESTED
Water contentKarl Fischer, coulometric, duplicate5.50%≤ 8.0PASS
Counter-ion (acetate)anion-exchange ion chromatography3.78%≤ 10.0PASS
Residual solvent + inorganic residueestimated — see /methods/residual-solvents-q3c/0.37%NOT TESTED
Endotoxinnot ordered on this submissionnot determined≤ 5.0NOT TESTED
Sterilitynot ordered on this submissionnot determinedno growthNOT TESTED
Appearancevisual, diffuse light, black and white backgroundOff-white lyophilised cake, intactwhite to off-white intact cakePASS
Reconstitution clarityreconstituted per submitter procedureClear, colourless, no visible particulatesclear, no visible particulatesPASS
pH (reconstituted, 1 mg/mL)potentiometric, two-point calibration5.672.5 – 8.0PASS
Retention time (observed)RP-HPLC, 25 min gradient5.90min6.71 ± 0.25 (normalised)FAIL
Peak asymmetry (10 % height)Ph. Eur. 2.2.461.290.80 – 1.50PASS
Plate count, main peakN = 5.54 (t/w½)²58,690≥ 20,000PASS
Resolution to nearest impurityvs deamidated Gln4 (+1 Da)1.84no index criterionREPORTED
Section 2

Chromatogram and peak integration

The figure is drawn from the table. It is not an independent illustration — if a number in the table changed, the trace would change with it.

0414082801242016560 Oxytocin5.90 min · 99.73 % 5.477.72 same trace · ×200 vertical expansion, baseline-corrected · main peak off-scale reporting threshold 0.05 % area 0481216202428 Retention time (min) Response (mAU)
Figure 1. RP-HPLC trace, 25 min gradient, 214 nm. Drawn from the peak-integration table below: peak positions are the reported retention times, peak areas are the reported area percentages, peak widths follow the reported plate count of 58,690, and skew follows the reported asymmetry factor of 1.29. Baseline noise is seeded from the report ID and is illustrative. The lower panel is the same trace at a stated vertical expansion with the slow gradient drift subtracted, because at full scale a 0.10 % impurity against a 99.73 % main peak is a line two pixels tall; the main peak runs off-scale there by design.
#PeakRTArea (µV·s)% areaΔ mass (Da)
1Oxytocin — main peak5.904,847,18799.73
2open-chain (reduced disulfide, +2 Da)5.474,8600.10+2.016
3oxytocin dimer (intermolecular disulfide)7.724,3740.09+1004.421
4Tyr sulfoxide artefact (+16 Da)5.671,9440.04+15.995
5deamidated Gln4 (+1 Da)5.721,9440.04+0.984
Closure check — purity + Σ impurities 100.00CLOSES

Reporting threshold 0.05 % area. Peaks below it are not reported and are therefore excluded from the total, which overstates purity by their sum — see peak integration.

Section 3

Mass spectrum and identity

Theoretical monoisotopic mass 1006.4365 Da, computed from C43H66N12O12S2. All ppm figures on this site are against monoisotopic values — see why that matters.

025507510025040055070085010001150 [M+1H]1+1007.44[M+2H]2+504.23 m/z Relative abundance (%)
Figure 2. ESI charge-state envelope, unit-resolution instrument. Sticks are placed at m/z = (M + z × 1.007276)/z for the observed neutral mass of 1006.4364 Da. Sodium (+21.982/z) and potassium (+37.956/z) adducts are shown where the parent charge state exceeds 25 % relative abundance; they are adducts of the same molecule and are not impurities.
ChargeTheoretical m/zObserved m/zΔ m/zRel. abundance (%)
3+336.4861336.4861-0.00009.2
2+504.2255504.2255-0.000057.0
1+1007.44381007.4437-0.0001100.0
Section 4

Gravimetric mass balance

Water, counter-ion, residual and total peptide content are exhaustive and mutually exclusive categories. They must account for 100 % of the cake mass. The index closes this to within 0.02 pp and the build asserts it.

Supplier certificate, as declared

The supplier's own figures for this lot, beside the independent measurement. Divergence is assessed against the batch median, not against this single report — see rule AX-R05.

FigureSupplier COAThis reportIndex note
Purity (% area)99.7399.73divergence -0.20 pp vs batch median
Total impurities (% area)0.270.27
Largest single impurity (% area)not stated0.10
Water (%)5.55.50
Counter-ion (%)not stated3.78
Peptide content (%w/w)not statednot determined
Endotoxin (EU/mg)4.09not determined
Method stated on certificateNOno method section — the result cannot be reconciled against any other
Section 5

Provenance and caveats

Where a sample came from changes the weight of the evidence. The index treats it as a first-class field.

Sample provenance
Buyer-sourced
Evidentiary weight
Strongest available evidence in this index.
Chain of custody
Retail purchase → submitter (ambient, 26 d) → laboratory (hand-carried, 5 d) → assay.
Assay date
13 Sep 2024
Submitted to the index
17 Sep 2024
Laboratory
Medutest, Ljubljana, Slovenia
Method basis
25 min gradient · unit-resolution MS
Verification status
IDENTITY CONSISTENT
Section 6

Annotations 4

Signed technical comments from index contributors. Not a discussion forum — annotations question or contextualise the result, and the data team replies where a reply is warranted.

aib_coupling
Hindered couplings are where a synthesis loses purity, and the deletion products they leave are the first thing to look for. On this record they are not there in quantity.
lambda_214
Detection wavelength does not affect a mass determination but it does affect the purity figure sitting next to it. Both are stated on this record, which is the point.
moq_reality
Retail purchases test the retail channel. Manufacturer-tier material bought at kilogram scale is a different product with different documentation, and this index sees very little of it.
ms_confirm
Retention alone never establishes identity. Retention plus mass does, and where the two disagree it is nearly always retention that is wrong for a documented reason.
Section 7

Related records

Same batch — lot HOMO-2405-U3389 n=14

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2024-004706HOMO-2405-U3389Janoshik2024-Q399.935.57104.1Buyer-sourced0PASS
AX-2024-004707HOMO-2405-U3389Janoshik2024-Q399.936.2894.4Buyer-sourced0PASS
AX-2024-004708HOMO-2405-U3389Janoshik2024-Q399.935.7490.4Buyer-sourced1FAIL
AX-2024-004709HOMO-2405-U3389Janoshik2024-Q399.935.12Buyer-sourced0PASS
AX-2024-004710HOMO-2405-U3389Janoshik2024-Q399.936.56101.4Buyer-sourced0PASS
AX-2024-004711HOMO-2405-U3389Nordanalyt2024-Q399.925.88104.0Buyer-sourced0PASS
AX-2024-004712HOMO-2405-U3389VendorInvestigate2024-Q399.736.2492.2Buyer-sourced0PASS
AX-2024-004713HOMO-2405-U3389VendorInvestigate2024-Q399.635.73102.3Laboratory-retained0PASS
AX-2024-004714HOMO-2405-U3389Nordanalyt2024-Q399.936.78105.5Supplier-supplied0PASS
AX-2024-004715HOMO-2405-U3389Janoshik2024-Q399.936.44Buyer-sourced0PASS
AX-2024-004716HOMO-2405-U3389Janoshik2024-Q399.935.57Buyer-sourced0FAIL
AX-2024-004717HOMO-2405-U3389Medutest2024-Q399.866.13Buyer-sourced0PASS
AX-2024-004718HOMO-2405-U3389Janoshik2024-Q399.936.05Buyer-sourced0PASS
Other recent Oxytocin records in the index.
ReportSupplierLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-041398QYBQYB-2601-V4429Medutest2026-Q299.604.7391.8Buyer-sourced0PASS
AX-2026-041400QYBQYB-2601-V4429Medutest2026-Q299.514.55Buyer-sourced1MARGINAL
AX-2026-041393QYBQYB-2601-V4429PeptideMeter2026-Q299.694.63Buyer-sourced0PASS
AX-2026-041395QYBQYB-2601-V4429Janoshik2026-Q299.934.24101.7Buyer-sourced0PASS
AX-2026-041950AMBAMB-2602-G4870Janoshik2026-Q299.823.21Buyer-sourced0PASS
AX-2026-041397QYBQYB-2601-V4429Janoshik2026-Q299.624.38102.8Buyer-sourced0PASS
AX-2026-041396QYBQYB-2601-V4429PeptideMeter2026-Q299.693.97Buyer-sourced0PASS
AX-2026-041394QYBQYB-2601-V4429Medutest2026-Q299.665.17Buyer-sourced0PASS
Other recent Homopeptide records in the index.
ReportCompoundLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042437RetatrutideHOMO-2605-V4718Nordanalyt2026-Q398.683.67102.8Buyer-sourced0PASS
AX-2026-042424SemaglutideHOMO-2604-X3806PeptideMeter2026-Q398.854.8792.0Buyer-sourced0PASS
AX-2026-042503IGF-1 LR3HOMO-2607-P5131Janoshik2026-Q399.365.99Buyer-sourced0PASS
AX-2026-042492SS-31 (Elamipretide)HOMO-2604-D5484Medutest2026-Q399.932.40Buyer-sourced0PASS
AX-2026-042491SS-31 (Elamipretide)HOMO-2604-D5484Kestrel2026-Q399.933.24Supplier-supplied0PASS
AX-2026-042473LL-37HOMO-2602-G5165Janoshik2026-Q399.015.00Buyer-sourced0PASS
AX-2026-042441RetatrutideHOMO-2605-V4718Medutest2026-Q398.934.22Buyer-sourced0PASS
AX-2026-042474LL-37HOMO-2602-G5165Janoshik2026-Q399.185.44Buyer-sourced0PASS