AX-2025-019511

5-Amino-1MQ · 50 mg declared · Guangzhou JEEP Biotechnology Co., Ltd. · lot JEEP-2412-M9286

FAILIDENTITY CONSISTENT
Report ID
AX-2025-019511
Compound
5-Amino-1MQ
Compound class
Non-peptide adjuncts
Supplier
JEEP — Guangzhou JEEP Biotechnology Co., Ltd.
Batch / lot
JEEP-2412-M9286
Declared label claim
50mg
Laboratory
Janoshik Analytical
Assay date
23 Apr 2025
Submitted
24 Apr 2025
Quarter
2025-Q2
Submitter
ig.beaulieu
Sample provenance
Buyer-sourced
Section 1

Results

Measured against the index acceptance specification for the non-peptide adjuncts class. A specification is a decision the index made; a result is a measurement. See where each threshold came from.

ParameterMethodResultUnitIndex spec
PurityRP-HPLC, 30 min gradient, 214 nm99.93% area≥ 98.0PASS
Total impuritiesRP-HPLC, sum of reported peaks0.07% area≤ 2.0PASS
Largest single impurityRP-HPLC — 5-nitro-1MQ (unreduced precursor, +30 Da)0.04% area≤ 1.00PASS
Identity — theoretical monoisotopic masscalculated from molecular formula285.9967DaNOT TESTED
Identity — observed monoisotopic massLC-MS, high resolution (ESI-TOF/QTOF)285.9937DaNOT TESTED
Mass accuracyobserved vs theoretical-10.6ppm± 10FAIL
Peptide contentnot ordered on this submissionnot determinedrequired to check label claimNOT TESTED
Water contentKarl Fischer, coulometric, duplicate5.05%≤ 10.0PASS
Counter-ion (iodide)anion-exchange ion chromatography37.76%≤ 12.0FAIL
Residual solvent + inorganic residueestimated — see /methods/residual-solvents-q3c/0.94%NOT TESTED
Endotoxinnot ordered on this submissionnot determined≤ 5.0NOT TESTED
Sterilitynot ordered on this submissionnot determinedno growthNOT TESTED
Appearancevisual, diffuse light, black and white backgroundWhite lyophilised cake, intactwhite to off-white intact cakePASS
Reconstitution clarityreconstituted per submitter procedureClear with a single visible fibreclear, no visible particulatesFAIL
pH (reconstituted, 1 mg/mL)potentiometric, two-point calibration5.462.5 – 8.0PASS
Retention time (observed)RP-HPLC, 30 min gradient5.42min5.42 ± 0.25 (normalised)PASS
Peak asymmetry (10 % height)Ph. Eur. 2.2.460.930.80 – 1.50PASS
Plate count, main peakN = 5.54 (t/w½)²81,669≥ 20,000PASS
Resolution to nearest impurityvs free iodide (counter-ion, by IC)0.00no index criterionREPORTED
Section 2

Chromatogram and peak integration

The figure is drawn from the table. It is not an independent illustration — if a number in the table changed, the trace would change with it.

06322126451896725289 5-Amino-1MQ5.42 min · 99.93 % 6.10 same trace · ×200 vertical expansion, baseline-corrected · main peak off-scale reporting threshold 0.05 % area 05101520253035 Retention time (min) Response (mAU)
Figure 1. RP-HPLC trace, 30 min gradient, 214 nm. Drawn from the peak-integration table below: peak positions are the reported retention times, peak areas are the reported area percentages, peak widths follow the reported plate count of 81,669, and skew follows the reported asymmetry factor of 0.93. Baseline noise is seeded from the report ID and is illustrative. The lower panel is the same trace at a stated vertical expansion with the slow gradient drift subtracted, because at full scale a 0.04 % impurity against a 99.93 % main peak is a line two pixels tall; the main peak runs off-scale there by design.
#PeakRTArea (µV·s)% areaΔ mass (Da)
15-Amino-1MQ — main peak5.424,856,90899.93
25-nitro-1MQ (unreduced precursor, +30 Da)6.101,9440.04+29.974
3free iodide (counter-ion, by IC)5.421,4580.030.000
Closure check — purity + Σ impurities 100.00CLOSES

Reporting threshold 0.05 % area. Peaks below it are not reported and are therefore excluded from the total, which overstates purity by their sum — see peak integration.

Section 3

Mass spectrum and identity

Theoretical monoisotopic mass 285.9967 Da, computed from C10H11N2+ · I−. All ppm figures on this site are against monoisotopic values — see why that matters.

0255075100250275300325350375 [M+1H]1+287.00 m/z Relative abundance (%)
Figure 2. ESI charge-state envelope, high-resolution instrument. Sticks are placed at m/z = (M + z × 1.007276)/z for the observed neutral mass of 285.9937 Da. Sodium (+21.982/z) and potassium (+37.956/z) adducts are shown where the parent charge state exceeds 25 % relative abundance; they are adducts of the same molecule and are not impurities.
ChargeTheoretical m/zObserved m/zΔ m/zRel. abundance (%)
1+287.0040287.0010-0.0030100.0
Section 4

Gravimetric mass balance

Water, counter-ion, residual and total peptide content are exhaustive and mutually exclusive categories. They must account for 100 % of the cake mass. The index closes this to within 0.02 pp and the build asserts it.

Supplier certificate, as declared

The supplier's own figures for this lot, beside the independent measurement. Divergence is assessed against the batch median, not against this single report — see rule AX-R05.

FigureSupplier COAThis reportIndex note
Purity (% area)99.9099.93divergence -0.03 pp vs batch median
Total impurities (% area)0.100.07
Largest single impurity (% area)0.070.04
Water (%)not stated5.05
Counter-ion (%)not stated37.76
Peptide content (%w/w)not statednot determined
Endotoxin (EU/mg)not statednot determined
Method stated on certificateYEScolumn, gradient and wavelength given
Section 5

Provenance and caveats

Where a sample came from changes the weight of the evidence. The index treats it as a first-class field.

Sample provenance
Buyer-sourced
Evidentiary weight
Strongest available evidence in this index.
Chain of custody
Retail purchase → submitter (−20 °C, 9 d) → laboratory (courier, dry ice, 8 d) → assay.
Assay date
23 Apr 2025
Submitted to the index
24 Apr 2025
Laboratory
Janoshik Analytical, Ostrava, Czechia
Method basis
30 min gradient · high-resolution MS
Verification status
IDENTITY CONSISTENT
Section 6

Annotations 4

Signed technical comments from index contributors. Not a discussion forum — annotations question or contextualise the result, and the data team replies where a reply is warranted.

fill_homogeneity
A high figure on one vial is a statement about one vial. The within-lot spread on the batch rollup is what tells you whether the bulk was well mixed, and it is the rarer and more useful number.
gradient_25
The gradient is on the record, which is the first thing that should be true of any purity figure and often is not. 30 min at Janoshik puts this comparison on a documented footing.
lot_traceability
One retail vial is one draw from a fill. Two or three from the same lot would separate fill variation from lot quality, and that is the cheapest useful experiment in this whole field.
label_claim
A retail vial is also a test of the label. Declared claim 50 mg against what the gravimetry recovers is a question only a purchased vial can answer honestly.
Section 7

Related records

Same batch — lot JEEP-2412-M9286 n=9

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2025-019504JEEP-2412-M9286Janoshik2025-Q299.846.11Buyer-sourced0FAIL
AX-2025-019505JEEP-2412-M9286PeptideMeter2025-Q299.785.44107.4Buyer-sourced0FAIL
AX-2025-019506JEEP-2412-M9286Janoshik2025-Q299.936.19100.8Buyer-sourced0FAIL
AX-2025-019507JEEP-2412-M9286Medutest2025-Q299.935.57101.4Supplier-supplied0FAIL
AX-2025-019508JEEP-2412-M9286Janoshik2025-Q299.765.07Buyer-sourced0FAIL
AX-2025-019509JEEP-2412-M9286PeptideMeter2025-Q299.355.87Buyer-sourced0FAIL
AX-2025-019510JEEP-2412-M9286PeptideMeter2025-Q299.936.04Buyer-sourced0FAIL
AX-2025-019512JEEP-2412-M9286Medutest2025-Q299.935.62104.7Buyer-sourced1FAIL
Other recent 5-Amino-1MQ records in the index.
ReportSupplierLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-043050KPKP-2511-W5347Janoshik2026-Q299.354.16105.5Buyer-sourced0FAIL
AX-2026-043044KPKP-2511-W5347PeptideMeter2026-Q299.614.1590.9Buyer-sourced0FAIL
AX-2026-043052KPKP-2511-W5347Medutest2026-Q299.454.23105.5Buyer-sourced1FAIL
AX-2026-043045KPKP-2511-W5347Janoshik2026-Q299.474.08Buyer-sourced0FAIL
AX-2026-043043KPKP-2511-W5347Janoshik2026-Q299.584.16101.1Buyer-sourced0FAIL
AX-2026-042869GGPepsGGPE-2512-M2434Kestrel2026-Q299.554.65103.9Buyer-sourced0FAIL
AX-2026-042868GGPepsGGPE-2512-M2434PeptideMeter2026-Q299.324.3988.5Buyer-sourced1FAIL
AX-2026-043047KPKP-2511-W5347Janoshik2026-Q299.684.17104.0Supplier-supplied0FAIL
Other recent JEEP records in the index.
ReportCompoundLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042177HexarelinJEEP-2608-Y9547Janoshik2026-Q399.903.42Buyer-sourced0PASS
AX-2026-042162SemaglutideJEEP-2607-X9091Janoshik2026-Q398.983.46Laboratory-retained1MARGINAL
AX-2026-042127TB-500JEEP-2602-V8614Medutest2026-Q399.683.55Buyer-sourced1FAIL
AX-2026-042123TB-500JEEP-2602-V8614PeptideMeter2026-Q399.933.80Buyer-sourced0PASS
AX-2026-042124TB-500JEEP-2602-V8614PeptideMeter2026-Q399.913.37104.9Buyer-sourced0PASS
AX-2026-042178HexarelinJEEP-2608-Y9547VendorInvestigate2026-Q399.684.03Buyer-sourced0PASS
AX-2026-042163SemaglutideJEEP-2607-X9091Nordanalyt2026-Q399.464.52Buyer-sourced1MARGINAL
AX-2026-042122TB-500JEEP-2602-V8614PeptideMeter2026-Q399.603.5897.0Buyer-sourced0PASS