5-Amino-1MQ by laboratory

The same compound measured by different laboratories on different gradients. Where the medians differ, the question is whether the material differed or the method did — and for 5-Amino-1MQ, with a class minimum gradient of 15 min, the method is a live explanation.

78 recordsmedian 99.45 %spec 98.0 %
LaboratoryReportsGradientMS resolutionMedian purity %Published bias (pp)Bias-adjustedMethod suitability
Janoshik3630minHigh99.49+0.0099.49ADEQUATE
PeptideMeter1328minHigh99.52-0.0599.57ADEQUATE
Medutest1325minUnit99.57+0.0899.49ADEQUATE
VendorInvestigate832minHigh98.92-0.1199.03ADEQUATE
Kestrel6 n=615minUnit99.70+0.2499.47ADEQUATE
Nordanalyt2 n=235minHigh99.46, 99.34+0.02ADEQUATE
97.898.298.699.099.399.7100.1JanoshikJanoshik: Q1 99.33, median 99.49, Q3 99.68, n=36outlier 98.19outlier 98.51outlier 98.71n=36PeptideMeterPeptideMeter: Q1 99.26, median 99.52, Q3 99.61, n=13outlier 98.55n=13MedutestMedutest: Q1 99.17, median 99.57, Q3 99.75, n=13outlier 98.27n=13VendorInvestigateVendorInvestigate: Q1 98.79, median 98.92, Q3 99.36, n=8n=8NordanalytNordanalyt: Q1 99.37, median 99.40, Q3 99.43, n=2n=2KestrelKestrel: Q1 99.53, median 99.70, Q3 99.91, n=6n=6 Purity (% area)
Figure. 5-Amino-1MQ purity by contributing laboratory. Any offset between rows has two possible sources — a real difference in the material each laboratory happened to receive, and the laboratory's own bias — and this figure cannot separate them. The paired analysis that can is on inter-lab agreement.