AssayIndex
Cross-section

LL-37 impurity profile

Every named species the index has reported alongside LL-37, how often, at what level, and what its presence implies about the synthesis or the handling. Reporting threshold is 0.05 % area; anything below it is invisible to this page and is silently added to the main peak.

17 recordsmedian 98.90 %spec 96.5 %
Median total impurities 1.10% n=17
Median largest single 0.68% n=17
Worst single observed 1.13% n=17
Named species observed 5 / 5 n=5
Named speciesΔRT (min)Δ mass (Da)Seen inMedian % areaMax % area
truncated (1-31)-1.68-709.401100%0.051.13
des-Leu1 (−113 Da)-0.38-113.084100%0.370.89
deamidated (+1 Da)-0.26+0.984100%0.160.33
trifluoroacetyl adduct (+96 Da)+0.84+95.982100%0.060.15
aggregated multimer+2.85mass-identical94%0.211.10

truncated (1-31)

Retention offset from the main peak
-1.68 min on RG-2
Mass difference
-709.401 Da
Reported in
17 of 17 records (100 %)
Level when reported
median 0.05 %, max 1.13 %
How it separates
Baseline-resolved on the reference method.

des-Leu1 (−113 Da)

Retention offset from the main peak
-0.38 min on RG-2
Mass difference
-113.084 Da
Reported in
17 of 17 records (100 %)
Level when reported
median 0.37 %, max 0.89 %
How it separates
Baseline-resolved on the reference method.

deamidated (+1 Da)

Retention offset from the main peak
-0.26 min on RG-2
Mass difference
+0.984 Da
Reported in
17 of 17 records (100 %)
Level when reported
median 0.16 %, max 0.33 %
How it separates
Close-eluting on the reference method, and merged with the main peak on a short gradient. The index prints the resolution figure for such a pair and applies no criterion to it — where the two are not separated, what matters is that the integration convention is stated.

trifluoroacetyl adduct (+96 Da)

Retention offset from the main peak
+0.84 min on RG-2
Mass difference
+95.982 Da
Reported in
17 of 17 records (100 %)
Level when reported
median 0.06 %, max 0.15 %
How it separates
Baseline-resolved on the reference method.

aggregated multimer

Retention offset from the main peak
+2.85 min on RG-2
Mass difference
mass-identical — chromatography only
Reported in
16 of 17 records (94 %)
Level when reported
median 0.21 %, max 1.10 %
How it separates
Chromatography only. A gradient long enough to separate it; no mass spectrometer of any resolution will.

Degradation pathways

Degradants are a handling finding rather than a synthesis finding. A lot that shipped clean and arrived oxidised tells you about the courier, not the chemist.

  • self-association into oligomers at physiological ionic strength
  • adsorptive loss on polypropylene — use glass