SS-31 (Elamipretide) by laboratory

The same compound measured by different laboratories on different gradients. Where the medians differ, the question is whether the material differed or the method did — and for SS-31 (Elamipretide), with a class minimum gradient of 15 min, the method is a live explanation.

124 recordsmedian 99.53 %spec 98.5 %
LaboratoryReportsGradientMS resolutionMedian purity %Published bias (pp)Bias-adjustedMethod suitability
Janoshik5130minHigh99.51+0.0099.51ADEQUATE
PeptideMeter2328minHigh99.47-0.0599.52ADEQUATE
Medutest2125minUnit99.69+0.0899.61ADEQUATE
Kestrel1315minUnit99.90+0.2499.66ADEQUATE
VendorInvestigate932minHigh98.63-0.1198.74ADEQUATE
Nordanalyt7 n=735minHigh99.73+0.0299.71ADEQUATE
96.997.598.098.599.199.6100.2JanoshikJanoshik: Q1 98.74, median 99.51, Q3 99.71, n=51n=51PeptideMeterPeptideMeter: Q1 98.88, median 99.47, Q3 99.73, n=23n=23MedutestMedutest: Q1 98.83, median 99.69, Q3 99.82, n=21outlier 97.16n=21VendorInvestigateVendorInvestigate: Q1 98.09, median 98.63, Q3 99.46, n=9n=9NordanalytNordanalyt: Q1 99.50, median 99.73, Q3 99.81, n=7n=7KestrelKestrel: Q1 99.10, median 99.90, Q3 99.93, n=13outlier 97.75n=13 Purity (% area)
Figure. SS-31 (Elamipretide) purity by contributing laboratory. Any offset between rows has two possible sources — a real difference in the material each laboratory happened to receive, and the laboratory's own bias — and this figure cannot separate them. The paired analysis that can is on inter-lab agreement.