Section 1

Results

Measured against the index acceptance specification for the tissue-repair and immunomodulatory peptides class. A specification is a decision the index made; a result is a measurement. See where each threshold came from.

ParameterMethodResultUnitIndex spec
PurityRP-HPLC, 35 min gradient, 214 nm99.55% area≥ 98.0PASS
Total impuritiesRP-HPLC, sum of reported peaks0.45% area≤ 2.0PASS
Largest single impurityRP-HPLC — [D-Val]-diastereomer0.25% area≤ 1.00PASS
Identitynot ordered on this submissionnot determinedrequired for VERIFIED statusNOT TESTED
Peptide contentnot ordered on this submissionnot determinedrequired to check label claimNOT TESTED
Water contentKarl Fischer, coulometric, duplicate5.64%≤ 10.0PASS
Counter-ion (acetate)anion-exchange ion chromatography5.41%≤ 12.0PASS
Residual solvent + inorganic residueestimated — see /methods/residual-solvents-q3c/0.93%NOT TESTED
Endotoxinkinetic chromogenic LAL1.05EU/mg≤ 5.0PASS
Sterilitynot ordered on this submissionnot determinedno growthNOT TESTED
Appearancevisual, diffuse light, black and white backgroundWhite lyophilised cake, intactwhite to off-white intact cakePASS
Reconstitution clarityreconstituted per submitter procedureClear after 90 s gentle swirling, no particulatesclear, no visible particulatesPASS
pH (reconstituted, 1 mg/mL)potentiometric, two-point calibration5.722.5 – 8.0PASS
Retention time (observed)RP-HPLC, 35 min gradient11.25min11.10 ± 0.25 (normalised)PASS
Peak asymmetry (10 % height)Ph. Eur. 2.2.460.960.80 – 1.50PASS
Plate count, main peakN = 5.54 (t/w½)²46,741≥ 20,000PASS
Resolution to nearest impurityvs [D-Val]-diastereomer0.77no index criterionREPORTED
Section 2

Chromatogram and peak integration

The figure is drawn from the table. It is not an independent illustration — if a number in the table changed, the trace would change with it.

02260451967799039 BPC-15711.25 min · 99.55 % 11.09 same trace · ×200 vertical expansion, baseline-corrected · main peak off-scale reporting threshold 0.05 % area 0510152025303540 Retention time (min) Response (mAU)
Figure 1. RP-HPLC trace, 35 min gradient, 214 nm. Drawn from the peak-integration table below: peak positions are the reported retention times, peak areas are the reported area percentages, peak widths follow the reported plate count of 46,741, and skew follows the reported asymmetry factor of 0.96. Baseline noise is seeded from the report ID and is illustrative. The lower panel is the same trace at a stated vertical expansion with the slow gradient drift subtracted, because at full scale a 0.25 % impurity against a 99.55 % main peak is a line two pixels tall; the main peak runs off-scale there by design.
#PeakRTArea (µV·s)% areaΔ mass (Da)
1BPC-157 — main peak11.254,838,43999.55
2[D-Val]-diastereomer11.0912,1510.250.000
3des-Gly truncate (−57 Da)10.856,3180.13-57.021
4BPC-157 arginine-salt artefact10.611,9440.040.000
5acetylated N-terminus (+42 Da)11.811,4580.03+42.011
Closure check — purity + Σ impurities 100.00CLOSES

Reporting threshold 0.05 % area. Peaks below it are not reported and are therefore excluded from the total, which overstates purity by their sum — see peak integration.

Section 4

Gravimetric mass balance

Water, counter-ion, residual and total peptide content are exhaustive and mutually exclusive categories. They must account for 100 % of the cake mass. The index closes this to within 0.02 pp and the build asserts it.

Supplier certificate, as declared

The supplier's own figures for this lot, beside the independent measurement. Divergence is assessed against the batch median, not against this single report — see rule AX-R05.

FigureSupplier COAThis reportIndex note
Purity (% area)99.1499.55divergence -0.41 pp vs batch median
Total impurities (% area)not stated0.45
Largest single impurity (% area)not stated0.25
Water (%)4.95.64
Counter-ion (%)not stated5.41
Peptide content (%w/w)90.3not determined
Endotoxin (EU/mg)not stated1.05
Method stated on certificateNOno method section — the result cannot be reconciled against any other
Section 5

Provenance and caveats

Where a sample came from changes the weight of the evidence. The index treats it as a first-class field.

Sample provenance
Buyer-sourced
Evidentiary weight
Strongest available evidence in this index.
Chain of custody
Retail purchase → third-party forwarder (8 d) → laboratory (courier, cold pack, 7 d) → assay.
Assay date
03 May 2026
Submitted to the index
18 May 2026
Laboratory
Nordanalyt Laboratories, Uppsala, Sweden
Method basis
35 min gradient · high-resolution MS
Verification status
PURITY ONLY
Section 6

Annotations 3

Signed technical comments from index contributors. Not a discussion forum — annotations question or contextualise the result, and the data team replies where a reply is warranted.

retest_date
Retail certificates in this index carry a retest date more often than they used to. It is the field that turns a release figure into a statement about the material over time.
gradient_25
Resolution scales with the square root of plate count and linearly with selectivity. Shallowing the gradient changes selectivity, which is the more effective of the two levers on a close pair.
plate_count
Efficiency is the lever that does not change selectivity, so it improves resolution without moving anything else. Longer column, smaller particle, or lower flow, in that order of cost.
Section 7

Related records

Same batch — lot TFC-2604-A9366 n=4

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042241TFC-2604-A9366PeptideMeter2026-Q299.466.07Buyer-sourced0PASS
AX-2026-042242TFC-2604-A9366PeptideMeter2026-Q299.626.06Buyer-sourced0PASS
AX-2026-042243TFC-2604-A9366Nordanalyt2026-Q299.545.61Laboratory-retained0PASS
Other recent BPC-157 records in the index.
ReportSupplierLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-043319CPCCPC-2602-W1610Medutest2026-Q398.822.88103.5Buyer-sourced2MARGINAL
AX-2026-041766TFSTFS-{CAT}-1156Janoshik2026-Q399.546.34Buyer-sourced1MARGINAL
AX-2026-043445BCHBCH-2603-G4350Nordanalyt2026-Q398.424.8793.6Buyer-sourced1MARGINAL
AX-2026-043440BCHBCH-2603-G4350VendorInvestigate2026-Q398.614.11Buyer-sourced1MARGINAL
AX-2026-043442BCHBCH-2603-G4350Janoshik2026-Q398.654.78Buyer-sourced1MARGINAL
AX-2026-043449BCHBCH-2603-G4350Kestrel2026-Q399.744.9291.3Supplier-supplied2MARGINAL
AX-2026-043448BCHBCH-2603-G4350Janoshik2026-Q398.554.69106.9Buyer-sourced1MARGINAL
AX-2026-043318CPCCPC-2602-W1610VendorInvestigate2026-Q398.473.12104.1Buyer-sourced1MARGINAL
Other recent TFC records in the index.
ReportCompoundLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042265Humanin (HNG)TFC-2605-U1300PeptideMeter2026-Q399.526.31Buyer-sourced0PASS
AX-2026-042222SS-31 (Elamipretide)TFC-2607-R9099VendorInvestigate2026-Q399.465.41102.6Buyer-sourced0PASS
AX-2026-042236Humanin (HNG)TFC-2608-S9940Nordanalyt2026-Q398.641.80Buyer-sourced0PASS
AX-2026-042282RetatrutideTFC-2608-J1156Janoshik2026-Q398.544.27Buyer-sourced0PASS
AX-2026-042226SS-31 (Elamipretide)TFC-2607-R9099Janoshik2026-Q399.625.47Buyer-sourced0PASS
AX-2026-042217SemaglutideTFC-2604-X9037PeptideMeter2026-Q398.954.6196.9Buyer-sourced0PASS
AX-2026-042339SermorelinTFC-2605-Y2176Kestrel2026-Q399.934.14Supplier-supplied1FAIL
AX-2026-042214SemaglutideTFC-2604-X9037Kestrel2026-Q399.874.66103.1Laboratory-retained0PASS