AssayIndex

AX-2026-042302

Epitalon · 20 mg declared · Tianjin Finder Chemical Co., Ltd. · lot TFC-2508-W1380

FAILMETHOD LIMITED
Report ID
AX-2026-042302
Compound
Epitalon
Compound class
Short bioregulator peptides
Supplier
TFC — Tianjin Finder Chemical Co., Ltd.
Batch / lot
TFC-2508-W1380
Declared label claim
20mg
Laboratory
Janoshik Analytical
Assay date
25 Jan 2026
Submitted
12 Feb 2026
Quarter
2026-Q1
Submitter
ghkcu_copper
Sample provenance
Buyer-sourced
Section 1

Results

Measured against the index acceptance specification for the short bioregulator peptides class. A specification is a decision the index made; a result is a measurement. See where each threshold came from.

ParameterMethodResultUnitIndex spec
PurityRP-HPLC, 30 min gradient, 214 nm98.77% area≥ 99.0FAIL
Total impuritiesRP-HPLC, sum of reported peaks1.23% area≤ 1.0FAIL
Largest single impurityRP-HPLC — β-Asp isomer0.71% area≤ 1.00PASS
Identity — theoretical monoisotopic masscalculated from molecular formula390.1387DaNOT TESTED
Identity — observed monoisotopic massLC-MS, high resolution (ESI-TOF/QTOF)390.1323DaNOT TESTED
Mass accuracyobserved vs theoretical-16.4ppm± 20PASS
Peptide contentnot ordered on this submissionnot determinedrequired to check label claimNOT TESTED
Water contentKarl Fischer, coulometric, duplicate6.31%≤ 12.0PASS
Counter-ion (acetate)anion-exchange ion chromatography9.29%≤ 10.0PASS
Residual solvent + inorganic residueestimated — see /methods/residual-solvents-q3c/0.93%NOT TESTED
Endotoxinnot ordered on this submissionnot determined≤ 5.0NOT TESTED
Sterilitynot ordered on this submissionnot determinedno growthNOT TESTED
Appearancevisual, diffuse light, black and white backgroundWhite lyophilised cake, intactwhite to off-white intact cakePASS
Reconstitution clarityreconstituted per submitter procedureClear, colourless, no visible particulatesclear, no visible particulatesPASS
pH (reconstituted, 1 mg/mL)potentiometric, two-point calibration4.832.5 – 8.0PASS
Retention time (observed)RP-HPLC, 30 min gradient4.65min4.62 ± 0.20 (normalised)PASS
Peak asymmetry (10 % height)Ph. Eur. 2.2.460.960.80 – 1.50PASS
Plate count, main peakN = 5.54 (t/w½)²60,117≥ 20,000PASS
Resolution to nearest impurityvs β-Asp isomer1.29no index criterionREPORTED
Section 2

Chromatogram and peak integration

The figure is drawn from the table. It is not an independent illustration — if a number in the table changed, the trace would change with it.

06152123031845524606 Epitalon4.65 min · 98.77 % 4.55 same trace · ×50 vertical expansion, baseline-corrected · main peak off-scale reporting threshold 0.05 % area 05101520253035 Retention time (min) Response (mAU)
Figure 1. RP-HPLC trace, 30 min gradient, 214 nm. Drawn from the peak-integration table below: peak positions are the reported retention times, peak areas are the reported area percentages, peak widths follow the reported plate count of 60,117, and skew follows the reported asymmetry factor of 0.96. Baseline noise is seeded from the report ID and is illustrative. The lower panel is the same trace at a stated vertical expansion with the slow gradient drift subtracted, because at full scale a 0.71 % impurity against a 98.77 % main peak is a line two pixels tall; the main peak runs off-scale there by design.
#PeakRTArea (µV·s)% areaΔ mass (Da)
1Epitalon — main peak4.654,800,52898.77
2β-Asp isomer4.5534,5080.710.000
3Ala-Glu-Asp tripeptide (−57 Da)4.4516,0390.33-57.021
4diketopiperazine (−18 Da)4.815,3460.11-18.011
5aspartimide (−18 Da)4.773,8880.08-18.011
Closure check — purity + Σ impurities 100.00CLOSES

Reporting threshold 0.05 % area. Peaks below it are not reported and are therefore excluded from the total, which overstates purity by their sum — see peak integration.

Section 3

Mass spectrum and identity

Theoretical monoisotopic mass 390.1387 Da, computed from C14H22N4O9. All ppm figures on this site are against monoisotopic values — see why that matters.

0255075100350375400425450475 [M+1H]1+391.14 m/z Relative abundance (%)
Figure 2. ESI charge-state envelope, high-resolution instrument. Sticks are placed at m/z = (M + z × 1.007276)/z for the observed neutral mass of 390.1323 Da. Sodium (+21.982/z) and potassium (+37.956/z) adducts are shown where the parent charge state exceeds 25 % relative abundance; they are adducts of the same molecule and are not impurities.
ChargeTheoretical m/zObserved m/zΔ m/zRel. abundance (%)
1+391.1460391.1396-0.0064100.0
Section 4

Gravimetric mass balance

Water, counter-ion, residual and total peptide content are exhaustive and mutually exclusive categories. They must account for 100 % of the cake mass. The index closes this to within 0.02 pp and the build asserts it.

Supplier certificate, as declared

The supplier's own figures for this lot, beside the independent measurement. Divergence is assessed against the batch median, not against this single report — see rule AX-R05.

FigureSupplier COAThis reportIndex note
Purity (% area)99.598.77divergence +0.49 pp vs batch median
Total impurities (% area)0.501.23
Largest single impurity (% area)0.360.71
Water (%)not stated6.31
Counter-ion (%)not stated9.29
Peptide content (%w/w)not statednot determined
Endotoxin (EU/mg)not statednot determined
Method stated on certificateYEScolumn, gradient and wavelength given
Section 5

Provenance and caveats

Where a sample came from changes the weight of the evidence. The index treats it as a first-class field.

Sample provenance
Buyer-sourced
Evidentiary weight
Strongest available evidence in this index.
Chain of custody
Retail purchase → submitter (ambient, 20 d) → laboratory (hand-carried, 3 d) → assay.
Assay date
25 Jan 2026
Submitted to the index
12 Feb 2026
Laboratory
Janoshik Analytical, Ostrava, Czechia
Method basis
30 min gradient · high-resolution MS
Verification status
METHOD LIMITED
Section 6

Annotations 3

Signed technical comments from index contributors. Not a discussion forum — annotations question or contextualise the result, and the data team replies where a reply is warranted.

deamidation_rate
Deamidation is time, temperature and sequence. If the deamidated species is prominent here, the manufacture date on lot TFC-2508-W1380 against the assay date above is the first arithmetic to do.
digit_preference
Rounding to the nearest half point loses about 0.25 pp of information in the worst case, which is inside the expanded uncertainty of the measurement anyway. The information loss is real and small.
method_absent
The more useful question than the rounding is whether the certificate states the method. Where it does, a rounded figure is comparable; where it does not, an unrounded one still is not.
Section 7

Related records

Same batch — lot TFC-2508-W1380 n=7

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042298TFC-2508-W1380Janoshik2026-Q199.017.0897.9Buyer-sourced1FAIL
AX-2026-042299TFC-2508-W1380Kestrel2026-Q198.846.70Buyer-sourced1FAIL
AX-2026-042300TFC-2508-W1380Janoshik2026-Q199.216.9099.7Buyer-sourced1MARGINAL
AX-2026-042301TFC-2508-W1380Medutest2026-Q198.826.6490.1Buyer-sourced1FAIL
AX-2026-042303TFC-2508-W1380Kestrel2026-Q199.356.9198.0Buyer-sourced1FAIL
AX-2026-042304TFC-2508-W1380Medutest2026-Q199.036.97104.4Buyer-sourced1MARGINAL
Other recent Epitalon records in the index.
ReportSupplierLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042008AMBAMB-2602-Y6433Janoshik2026-Q399.308.75105.1Buyer-sourced0PASS
AX-2026-042009AMBAMB-2602-Y6433VendorInvestigate2026-Q399.168.62102.3Buyer-sourced0FAIL
AX-2026-042010AMBAMB-2602-Y6433Janoshik2026-Q399.098.20103.4Laboratory-retained0FAIL
AX-2026-042007AMBAMB-2602-Y6433Janoshik2026-Q399.428.6293.8Buyer-sourced0FAIL
AX-2026-042468HomopeptideHOMO-2512-A5108Janoshik2026-Q299.937.81100.8Buyer-sourced0PASS
AX-2026-042463HomopeptideHOMO-2512-A5108Janoshik2026-Q299.937.6893.2Buyer-sourced0FAIL
AX-2026-042041HYBHYB-2603-U7133Janoshik2026-Q299.656.8295.3Buyer-sourced0FAIL
AX-2026-042470HomopeptideHOMO-2512-A5108Janoshik2026-Q299.637.99Buyer-sourced0FAIL
Other recent TFC records in the index.
ReportCompoundLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042265Humanin (HNG)TFC-2605-U1300PeptideMeter2026-Q399.526.31Buyer-sourced0PASS
AX-2026-042222SS-31 (Elamipretide)TFC-2607-R9099VendorInvestigate2026-Q399.465.41102.6Buyer-sourced0PASS
AX-2026-042236Humanin (HNG)TFC-2608-S9940Nordanalyt2026-Q398.641.80Buyer-sourced0PASS
AX-2026-042282RetatrutideTFC-2608-J1156Janoshik2026-Q398.544.27Buyer-sourced0PASS
AX-2026-042226SS-31 (Elamipretide)TFC-2607-R9099Janoshik2026-Q399.625.47Buyer-sourced0PASS
AX-2026-042217SemaglutideTFC-2604-X9037PeptideMeter2026-Q398.954.6196.9Buyer-sourced0PASS
AX-2026-042339SermorelinTFC-2605-Y2176Kestrel2026-Q399.934.14Supplier-supplied1FAIL
AX-2026-042214SemaglutideTFC-2604-X9037Kestrel2026-Q399.874.66103.1Laboratory-retained0PASS